MA601 / MA801 Error Messages

Page Error Message Reason Solution
Startup / Loading The product can not boot, please contact the dealer. Device is unable to turn on normally. It is possible electronic components or cables are malfunctioning Contact manufacturer for solutions
Main Page Please confirm cable connection to device. UA-010 Transfer Key Cable to Charder Insight has been unplugged Plug in UA-010 USB Transfer Key Cable
Weight Measurement Communication error, please restart device. Software error Restart device. If problem continues to occur without improvement, contact manufacturer for solutions
Preparing for scan The measurement has stopped working
Measuring Error
Subject steps off from platform before measurement is completed Return to Main Page, and try measurement again
Preparing for scan The measurement has stopped working
Insufficient electrode contact
Device is unable to successfully measure subject's bioelectrical impedance. It may be because the subject's hands and/or feet are not correctly contacting electrodes, or the surface of their skin is overly dry Make sure subject has removed their shoes and socks. If needed, clean their hands and feet, and check to make sure they are contacting electrodes properly
BIA Measurement The measurement has stopped working
Measuring Error
Subject steps off from platform before measurement is completed Return to Main Page, and try measurement again
BIA Measurement The measurement has stopped working
Insufficient electrode contact
During BIA scan, device has detected abnormality in electrode contact between hands/feet. Subject may have let go of hand electrodes, or moved their foot position Try measurement again
BIA Measurement Communication error, please restart device. Software error Restart device. If problem continues to occur without improvement, contact manufacturer for solutions
BIA Measurement The measurement has stopped working
Timeout
Subject's posture may have changed too much during measurement to the extent that device is unable to complete measurement Try measurement again
Measurement Result Please confirm cable connection to device. UA-010 Transfer Key Cable to Charder Insight has been unplugged Plug in UA-010 USB Transfer Key Cable, and measure again
Measurement Result Data transfer failure. Timeout during data transfer Try measurement again. If problem continues to occur without improvement, contact manufacturer for solutions
Measurement Result The history data is about to reach limit, please delete some data in Data manager. Device is close to maximum measurement storage limit Go to Data Manager and delete measurement results that are not needed anymore
Measurement Result Printing error
Unable to communicate with printer
Try the following:
Turn the printer off and then on again.
Make sure all cables are securely connected or make sure your wireless connection is working.
Device is unable to connect to printer
1. Confirm printer has PCL5 or above driver compatibility
2. Confirm device is connected to printer correctly using USB cable
3. Confirm printer drivers are installed on device